Table 3.

Reproducibility of the five independent potentiometric arrayed glucose biosensors based on Nafion-GOx/GO/AZO over glucose concentrations ranging from 2 to 10 mM (number of sensors, N = 5).

Number of Sensors Glucose Concentration
(mM)
Response Voltage
(Mean ± SD, mV)
Average Sensitivity (mV/mM) Linearity
1 2 146.64 ± 3.77 15.42 0.996
4 121.10 ± 4.36
6 91.54 ± 3.68
8 50.20 ± 4.16
10 27.88 ± 6.36
2 2 145.13 ± 4.57 15.14 0.997
4 123.12 ± 4.80
6 90.31 ± 5.47
8 53.49 ± 6.38
10 28.50 ± 5.71
3 2 142.87 ± 5.46 15.67 0.994
4 125.10 ± 4.92
6 89.35 ± 5.16
8 51.34 ± 6.01
10 23.07 ± 6.38
4 2 145.25 ± 4.25 15.38 0.996
4 124.69 ± 5.28
6 92.10 ± 6.16
8 53.46 ± 5.74
10 27.03 ± 5.95
5 2 144.51 ± 6.34 15.10 0.997
4 119.39 ± 4.31
6 89.37 ± 5.84
8 50.86 ± 3.17
10 27.79 ± 5.06