Table 3.
Reproducibility of the five independent potentiometric arrayed glucose biosensors based on Nafion-GOx/GO/AZO over glucose concentrations ranging from 2 to 10 mM (number of sensors, N = 5).
Number of Sensors | Glucose Concentration (mM) |
Response Voltage (Mean ± SD, mV) |
Average Sensitivity (mV/mM) | Linearity |
---|---|---|---|---|
1 | 2 | 146.64 ± 3.77 | 15.42 | 0.996 |
4 | 121.10 ± 4.36 | |||
6 | 91.54 ± 3.68 | |||
8 | 50.20 ± 4.16 | |||
10 | 27.88 ± 6.36 | |||
2 | 2 | 145.13 ± 4.57 | 15.14 | 0.997 |
4 | 123.12 ± 4.80 | |||
6 | 90.31 ± 5.47 | |||
8 | 53.49 ± 6.38 | |||
10 | 28.50 ± 5.71 | |||
3 | 2 | 142.87 ± 5.46 | 15.67 | 0.994 |
4 | 125.10 ± 4.92 | |||
6 | 89.35 ± 5.16 | |||
8 | 51.34 ± 6.01 | |||
10 | 23.07 ± 6.38 | |||
4 | 2 | 145.25 ± 4.25 | 15.38 | 0.996 |
4 | 124.69 ± 5.28 | |||
6 | 92.10 ± 6.16 | |||
8 | 53.46 ± 5.74 | |||
10 | 27.03 ± 5.95 | |||
5 | 2 | 144.51 ± 6.34 | 15.10 | 0.997 |
4 | 119.39 ± 4.31 | |||
6 | 89.37 ± 5.84 | |||
8 | 50.86 ± 3.17 | |||
10 | 27.79 ± 5.06 |