Table 4.

Potentiometric glucose biosensors based on different electrodes [8,9,10,55,56,57,58].

Electrode Linear Range Sensitivity Linearity Response Time LOD Ref.
Nafion-GOx/GO/AZO/Ag 2 mM to 10 mM 15.44 mV/mM 0.997 26 s 1.89 mM This study
BSA-Nafion-GOx/ZnO NWs/Ag 0.5–1000 μM 35 mV/decade N/A 1–4 s N/A [9] 2010
BSA-GOx/ZnO NFs/Al 500 nM to 10 mM 65.2 mV/decade 0.990 4 s N/A [10] 2010
GOx/ZnO NRs/Ag 1 µM to 10mM 2.51 mV/decade 0.980 N/A N/A [11] 2018
Fe3O4-GOx-Ppy/MGCE 0.5 μM t 34 mM 19.4 mV/decade 0.998 6 s 0.3 μM [56] 2014
PAPBAOT/GCE 5–50 mM 0.58 mV/mM 0.992 200 s 0.5 mM [57] 2013
AgNPs-GOx/Ag-ISE 0.1–3 mM 8.62 mV/decade N/A N/A 10 μM [58] 2009
MIP-based on MAA 0.02–7 mM 43.7 mV/mM 0.980 N/A N/A [59] 2017

Note: GOx: glucose oxidase; GO: graphene oxide; AZO: aluminum-doped zinc oxide; ZnO: zinc oxide; NWs: nanowires; BSA: bovine serum albumin; NFs: nanoflakes; Ppy: polypyrrole; MGCE: magnetic glassy carbon electrode; GCE: glassy carbon electrode; PAPBAOT: poly (3-aminophenyl boronic acid-co-3-octylthiophene); MIP: molecularly imprinted polymer; MAA: methacrylic acid; NRs: nanorods.