Table 1.
Valve characteristics and features.
ACURATE neo (Boston Scientific Corporation) | SAPIEN 3 (Edwards Lifesciences) |
---|---|
![]() |
![]() |
Supra-annular | Intra-annular |
Porcine pericardial leaflet tissue | Bovine pericardial leaflet tissue |
Self-expanding, deployment in a top-down mechanism of nitinol frame. | Balloon-expandable cobalt-chromium frame |
Transfemoral sheath size (valve size) | |
18-French for all devices: Small (23 mm), Medium (25 mm), Large (27 mm). | Ready for ultra-low profile: 14 F (20, 23, 26 mm); 16 F (29 mm), 18 F (20, 23, 26 mm), 21 F (29 mm) |
Special features | |
-Upper and lower crown; -Three stabilization arches; -Outer and inner pericardial skirt. |
-Outer sealing and inner skirt at the inflow |