Table 3. The impact of the initial starting values of the refractive index for OCT (nOCT) and THz (nTHz) on the outcome of the algorithm for the case example of pair combination [A,E]. A percentage difference was calculated for each value compared to the measured value. Note that the measured values for d1 and d2 are 3.05 mm and 1.27 mm respectively for regions A and E. The measured refractive index for nOCT is 1.43, and measured refractive index for THz, nTHz, is 1.53.
Initial starting values | Difference between algorithm and measured depth (%) | Difference between algorithm and measured refractive index (%) | |||
---|---|---|---|---|---|
nOCT initial | nTHz initial | d 1 | d 2 | n OCT | n THz |
1.300 | 1.500 | 7.7 | 3.2 | −5.9 | −6.7 |
1.300 | 1.600 | 7.7 | 3.2 | −5.9 | −6.7 |
1.300 | 1.700 | 7.7 | 3.2 | −5.9 | −6.7 |
1.400 | 1.500 | 3.9 | 1.6 | −2.5 | −2.7 |
1.400 | 1.600 | 3.9 | 1.6 | −2.5 | −2.7 |
1.400 | 1.700 | 3.9 | 1.6 | −2.5 | −2.7 |
1.450 | 1.500 | 2.2 | 0.9 | −0.9 | −0.8 |
1.450 | 1.550 | 2.2 | 0.9 | −0.9 | −0.8 |
1.450 | 1.600 | 2.2 | 0.9 | −0.9 | −0.8 |
1.450 | 1.700 | 2.2 | 0.9 | −0.9 | −0.8 |
1.500 | 1.400 | 0.5 | 0.2 | 0.8 | 1.0 |
1.500 | 1.500 | 0.5 | 0.2 | 0.8 | 1.0 |
1.500 | 1.600 | 0.5 | 0.2 | 0.8 | 1.0 |
1.500 | 1.700 | 0.5 | 0.2 | 0.8 | 1.0 |