Skip to main content
. 2020 Feb 12;12(2):150. doi: 10.3390/pharmaceutics12020150

Table 4.

Trueness, precision and accuracy results for each concentration level in the validation data sets.

True PRX Concentration (% w/w) Mean Predicted PRX Concentration (% w/w) Trueness Precision Accuracy
Relative Bias (%) Recovery (%) Repeatability (RSD%) Intermediate Precision (RSD%) Relative β-Expectation Tolerance Limits (%)
11.66 11.65 −0.12 99.88 0.55 0.84 [−2.92; 2.68]
13.45 13.51 0.47 100.47 0.54 0.95 [−3.51; 4.45]
15.45 15.40 −0.29 99.71 0.65 1.05 [−4.05; 3.47]
17.50 17.48 −0.14 99.86 0.80 1.09 [−3.13; 2.85]