Skip to main content
. 2020 Mar 17;7:18. doi: 10.1186/s40658-020-0284-5

Table 1.

Percent FWHM measured on CzT and Anger cameras for 6 energy peaks. Results are expressed in mean (SD)

Radioisotope 201Tl 99mTc 123I 111In
Energy peak (keV) 70 167 140 159 171 245
% FWHM Anger camera 14.08 (0.19) 11.98 (1.66) 9.21 (0.07) 10.24 (0.79) 10.78 (0.03) 9.03 (0.35)
% FWHM CzT camera 5.77 (1.30) 4.67 (1.13) 5.46 (0.59) 5.33 (0.62) 6.27 (1.08) 4.5 (0.75)