Table 2.
Microbeam detection method | |||
---|---|---|---|
vM1212 detector | Zeiss Axio Imager 2 (+ EBT3 film) | Epson Expression 10000XL (+ EBT3 film) | |
Advantages |
Real time measurement and analysis Short exposure is sufficient to obtain accurate profile information |
Highest spatial resolution No dose rate dependence49 |
Lower cost Established clinical workflow No dose rate dependence49 |
Disadvantages |
Limited life expectancy due to cumulative radiation damage Spatial resolution limited by 50 µm pixel pitch |
24 hours self‐development Complex and time consuming analysis process |
24 hours self‐development Poorest spatial resolution ad hence limited suitability for microbeam applications |
Higher price | Necessity to establish procedures and workflow for wider uptake of this method | Software licensing costs |