Pentacam measurements versus ultrasound pachymetry measurements (μm)

Parameters Bias SD 95% Limits of Agreement (μm)
Upper Lower Width
Preoperative corneal pachymetry
 Corneal Vertex −1.59 7.80 13.70 −16.88 30.58
 Pupil Center −1.09 7.64 13.87 −16.06 29.93
Postoperative corneal pachymetry
 Corneal Vertex 2.07 6.46 14.73 −10.60 25.34
 Pupil Center 2.15 6.51 14.91 −10.61 25.52