Published in final edited form as: IEEE Trans Biomed Eng. 2019 May 8;67(2):471–481. doi: 10.1109/TBME.2019.2915526

Table 3.

Comparison of PID and rule-based fluid resuscitation controllers

PID Rule-Based P-value
Dynamic Performance Response Time [0 to 90%] (min) 25.73 ± 14.53 27.56 ± 20.63 P=0.850 (T)*
Settling Time [5%] (min) 39.08 ± 14.88 40.65 ± 17.01 P=0.562 (T)
Relative Overshoot (%) 1.90 ± 5.52% 3.08 ± 4.23% P=0.636 (T)
Measurement Errors SSE (%) 2.84 ± 6.00 9.79 ± 4.37 P<0.001 (T)*
RMSE (%) 9.17 ± 4.70 13.01 ± 5.95 P=0.002 (T)*
PE (%) −6.20 ± 8.08 −12.23 ± 7.16 P<0.001 (WSR)**
Bias (%) −4.59 ± 6.21 −11.10 ± 5.33 P<0.001 (T)*
Inaccuracy (%) 6.61 ± 3.76 11.29 ± 4.89 P=0.003 (T)*
Wobble (%) 1.15 ± 0.89 1.51 ± 0.94 P=0.239 (T)*
Divergence (% h−1) −9.09 ± 8.01 −9.38 ± 7.44 P=0.885 (T)**
Infusion and MAP Response Infusion Rate (mL/min) 17.46 ± 9.06 10.51 ± 7.51 P<0.001 (WSR)
Total Infusion (mL) 1048 ± 472 631 ± 312 P<0.001 (T)
MAP Response (mmHg) 75.04 ± 6.46 70.22 ± 5.73 P<0.001 (WSR)
*

Paired T-test

**

WSR: Wilcoxon Signed-Rank Test