Table 2.

Diameter of the computed radiation isocentre for each phantom for TrueBeam with aS1200 EPID.

Series number Component PTW Isoball Aquilab Isocentre Phanton Varian Isocentre Cube
1 Collimator 0.70 0.73 0.72
Gantry 0.41 0.33 0.65
Couch 0.72 0.70 1.92
2 Collimator 0.77 0.72 0.75
Gantry 0.18 0.24 0.70
Couch 1.10 0.41 1.47
3 Collimator 0.80 0.72 0.68
Gantry 0.30 0.29 0.78
Couch 1.60 1.25 1.14
4 Collimator 0.72 0.69 0.68
Gantry 0.28 0.31 0.80
Couch 0.70 0.82 0.69
5 Collimator 0.89 0.74 0.73
Gantry 0.33 0.29 0.79
Couch 0.38 0.67 1.14
6 Collimator 0.72 0.70 0.73
Gantry 0.31 0.39 0.88
Couch 0.39 1.32 0.81