Skip to main content
. 2020 Feb 21;16(8):1363–1375. doi: 10.7150/ijbs.41395

Table 4.

Accuracy and precision at ROI level on QCs at two concentrations of olaparib

Spotted pmol/spot Nominal pmol/mm2 Measured pmol/mm2 MEAN SD CV (%) Accuracy (%) Accuracy MEAN (%)
7 4.5 4.6 4.8 0.2 3.4 2.6 27.2
3.1 4.7 52.3
3.9 5.0 26.6
3 2.0 1.8 1.8 0.3 14.5 12.0 12.2
1.8 1.5 18.6
1.9 2.0 6.2