Skip to main content
. 2020 Feb 29;20(5):1346. doi: 10.3390/s20051346

Figure 7.

Figure 7

(a) Reflection spectra of device under different temperatures; (b) resonant frequency as function of temperature, showing temperature coefficient of −4630 ppm/K frequency, calculated from gradient; (c) frequency at each temperature test point for 50 times; (d) frequency shift under cyclic temperature scanning, showing good consistency with no visible hysteresis.