Table 7.
Comparisons of the accuracy requirements for blood glucose meters, as dictated by the ISO 15197:2013, and iCGM, as required by the US FDA.
| ISO 15197:2013 | FDA iCGM | |
|---|---|---|
| Overall Accuracy | ≥95% within ± 15 mg/dL for glucose concentrations <100 mg/dL ≥95% within ± 15% for glucose concentrations >100 mg/dL |
>87% within ± 20% |
| ACCURACY REQUIREMENTS BY GLUCOSE CONCENTRATION | ||
| <70 mg/dL | n/a | >85% within ± 15 mg/dL >98% within ± 40 mg/dL |
| 70–180/dL | n/a | >70% within ± 15% >99% within ± 40% |
| >180 mg/dL | n/a | >80% within ± 15% >99% within ± 40% |
| ADDITIONAL REQUIREMENTS | ||
| ≥99% within consensus error grid zones A and B | ≤1% of glucose rates of change >1 mg/dL/min if true rate of change is < −2 mg/dl/min ≤ 1% of glucose rates of change < −1 mg/dL/min if true rate of change is >2 mg/dL/min |
|
Adapted from Freckmann et al. (56).