Skip to main content
. 2020 Mar 24;10:5261. doi: 10.1038/s41598-020-62061-4

Table 2.

Data accuracy between polysomnography (PSG) and impulse-radio ultra-wideband (IR-UWB) radar according to the severity of obstructive sleep apnea (OSA).

PSG (AHI) IR-UWB radar (ABI) Recall
Control Mild OSA Moderate OSA Severe OSA
Control 17 0 0 0 1
Mild OSA 6 19 3 0 0.68
Moderate OSA 0 5 11 0 0.69
Severe OSA 0 0 0 33 1
Precision 0.74 0.79 0.78 1 0.85

The severity of OSA was categorized as follows: control (ABI or AHI < 5), mild (5 ≤ ABI or AHI < 15), moderate (15 ≤ ABI or AHI < 30), and severe (ABI or AHI ≥ 30).

ABI, Abnormal Breathing Index; AHI, Apnea–hypopnea Index.

The value of the overall agreement is 0.85.