Skip to main content
. 2013 Apr 4;134(3):333–340. doi: 10.1016/j.exppara.2013.03.031

Table 3.

Sensitivity and specificity of the LAMP-Tube scanner, LAMP-Thermocycler and RDTs compared to microscopy.

Microscopy (n) RDTs
LAMP-Thermocycler
LAMP-Tube scanner
Positive Negative Positive Negative Positive Negative
Positive (60) 52 8 58 2 57 3
Negative (30) 5 25 2 28 2 28



Sensitivity (%) 86.7 96.7 95.0
Specificity (%) 83.3 93.3 93.3