Major Sources of Contamination

Source of Contamination Reference(s)
Ineffective cleaning
 Inadequate cleaning 17, 71
 Damaged internal channel 18, 19
 Poorly mated internal components 74
 Reusable suction valve 20
 Suction channel 20, 21
 Biopsy port 71
Accessories
 Sample collection tubing 22
 Reused stopcocks for BAL fluid aspiration 23
Contaminated reprocessing equipment
 Automated washer 69
 Rinsing tank 24
 Tubing 25
 Filter 26
 Biofilm in reprocessor 27, 69
 Cleaning brushes 28
Instilled solutions
 Topical anesthesia (cocaine) 29
 Green dye (additive to anesthetic) 30
 Atomizer 31
Disinfectant
 Inadequate activity 32
 Incorrect disinfectant concentration dispensed by automated reprocessor 73
 Contaminated glutaraldehyde 23, 33
Improper connector to reprocessor 70
Recontamination after disinfection
 Rinsing tap water (hospital supply) 34, 35, 36, 37
 Contaminated water filters 38, 39
 Reuse of “sterile water” for rinsing 40
 Reassembly of valves prior to storage 41
Storage in coiled position/in cases: 42