. 2020 Mar 10;70(693):e236–e244. doi: 10.3399/bjgp20X708845

Table 3.

Diagnostic accuracy for fever defined as ≥38°C by the electronic axillary thermometer

Thermofocus Firhealth Tympanic
Sensitivity, % (95% CI) 66.7 (38.4 to 88.2) 12.5 (1.6 to 38.3) 62.6 (35.4 to 84.8)
Specificity, % (95% CI) 98.0 (96.0 to 99.2) 99.4 (98.0 to 99.9) 96.0 (93.4 to 97.8)
Positive predictive value, % (95% CI) 58.8 (32.9 to 81.6) 50.0 (6.8 to 93.2) 41.7 (22.1 to 63.4)
Negative predictive value, % (95% CI) 98.6 (96.7 to 99.5) 96.2 (93.7 to 97.9) 98.2 (96.2 to 99.4)
Likelihood ratio + (95% CI) 33.9 (15.0 to 76.7) 22.4 (3.4 to 148.8) 15.6 (8.2 to 29.5)
Likelihood ratio – (95% CI) 0.34 (0.17 to 0.70) 0.88 (0.73 to 1.06) 0.39 (0.21 to 0.74)
Absolute numbers (TP, FP, FN, TN) 10, 5, 7, 349 2, 14, 2, 356 10, 6, 14, 335

FN = false negative. FP = false positive. TN = true negative. TP = true positive.