Table 3.

ROC values of new Corvis ST parameters in identifying KE eyes from LASIK eyes.

Parameter AUC(95 %CI) P Youden index Cut off Sensitivity (%) Specificity (%)
Max Inverse Radius 0.850 (0.714, 0.938) <0.001 0.565 >0.227 82.61 73.91
DA Ratio Max [2 mm] 0.847 (0.710, 0.936) <0.001 0.609 >6.733 65.22 95.65
Pachy Slope 0.664 (0.509, 0.796) 0.055 0.435 >128.391 56.52 86.96
DA Ratio Max [1 mm] 0.750 (0.601, 0.866) 0.001 0.565 >1.779 78.26 78.26
ARTh 0.686 (0.533, 0.815) 0.021 0.391 ≤76.134 39.13 100.00
Integrated Radius 0.885 (0.756, 0.960) <0.001 0.696 >12.473 86.96 82.61
SP-A1 0.851 (0.715, 0.938) <0.001 0.652 ≤62.422 78.26 86.96
CBI 0.543 (0.390, 0.691) 0.148 0.087 >0.999 100.00 8.70

ARTh, Ambrosio’s relational thickness horizontal; AUC(95 %CI): area under the ROC curve (95 % Confidence interval); CBI, Corvis biomechanical index; DA Ratio Max [2 mm], deformation amplitude Ratio Max [2 mm]; DA Ratio Max [1 mm], deformation amplitude Ratio Max [1 mm]; SP-A1, stiffness parameter at first applanation.