Table 2.
Results of the reproducibility evaluation for the eight analyzed panels. Test performed during 5 days with two repeats for each panel.
Panels | Ct mean | % CV | |
---|---|---|---|
FLUAV-H1pdm09 | High titer | 24.2 | 3.3 |
Low titer | 34.2 | 1.8 | |
FLUAV-H3 | High titer | 27.5 | 3.9 |
Low titer | 33.9 | 7.2 | |
FLUBV | High titer | 26.6 | 6.5 |
Low titer | 34.4 | 8.9 | |
HRSV | High titer | 30.7 | 11.2 |
Low titer | 36.2 | 5.6 |
Abbreviations: CV = variance coefficient.