Skip to main content
. 2004 Jan 1;50(1):67–72. doi: 10.1373/clinchem.2003.023663

Table 3.

Mean (SD) differences in Ct values between quantitative assays for ORF1b and N gene.

Calibrator (n = 14) Infected cells (n = 9) NPA (n = 18) Stool (n = 11)
ΔCt1 0.29 (1.03) 1.32 (0.67)2 0.34 (0.79)c 0.14 (0.79)d
1

ΔCt, Ct value of a sample in the ORF 1b assay − Ct value of the same sample in the N gene assay.

2

ΔCt statistically different from that of the calibrator (Student t-test, P <0.02; 95% confidence interval, 0.39–1.67).

c,d ΔCt not statistically different from that of the calibrator (Student t-test): cP = 0.89; dP = 0.64.