Skip to main content
. 2005 Jul 12;35(1):51–58. doi: 10.1016/j.jcv.2005.05.005

Table 1.

Panel composition and results of pre-distribution testing at reference laboratories

Virus Strong positive (A)
Positive (B)
Low positive (C)
Dilution Ct Dilution Ct Dilution Ct
Influenza virus Aa 10E−3 27 10E−5 34.5 10E−6 37.9
Influenza virus Ba 10E−3 25.3 10E−5 33.4 10E−6 37.3



Parainfluenza virus 1b 10E−3 35.8 10E−4 38.9 10E−5 44.9
Parainfluenza virus 3b 10E−3 28.1 10E−4 33.5 10E−5 35.8



Respiratory syncytial virusa 10E−3 27.4 10E−5 35.7 10E−6 38
Human metapneumovirusa 10E−1 24.5 10E−3 32.5 10E−4 37.6



Human rhinovirus 16a 10E−3 23.8 10E−5 31.1 10E−6 36.2
Human rhinovirus 72b 10E−3 24.2 10E−5 30.1 10E−6 34.7
Human rhinovirus 90a 10E−3 26.1 10E−5 32.8 10E−6 36.5



Adenovirus 4b 10E−2 31.8 10E−4 37.7 10E−5 41.8
Adenovirus 7b 10E−2 30.8 10E−4 38.2 10E−5 41.0



Human coronavirus 229Eb 10E−3 27.6 10E−5 34 10E−6 36.6
Human coronavirus OC43a 10E0 30.7 10E−1 33.5 10E−2 37.2

Ct: cycle threshold values.

a

Ct values shown are a mean of results from two laboratories.

b

Ct values shown are a mean of results from a single laboratory.