. 2020 Feb 19;61(4):523–536. doi: 10.1194/jlr.P119000325

TABLE 1.

Calibration curve for the acyl-CoA species

Linearity Precision [CV(%)]a Accuracy (%)
Compound Range (pmol) Weight Slope Intercept r2 QC-Lb QC-Mb QC-Hb QC-Lb QC-Mb QC-Hb
17:0-CoA/(D31)-16:0-CoA 0.05–100 1/x2 0.061 −0.0024 0.991 Intra-day (n = 12) 10.6 6.3 3.7 9.5 5.2 5.8
Inter-day (n = 4+4+4) 21.2 10.3 6.3 21.2 11.8 9.2
a

Precision was calculated as the coefficient of variation (CV).

b

Three samples with 0.050, 1.0, and 10 pmol of 17:0-CoA were mixed with 10 pmol of (D31)-16:0-CoA per injection and were analyzed as QC compounds QC-L (low), QC-M (middle), and QC-H (high), respectively.