Table 3.
Analyses of correlation (pearson) and agreement (intraclass correlations) among three different devices and axillary temperature measurements.
| Correlation coefficient |
Agreement |
||||||
|---|---|---|---|---|---|---|---|
| Pearson (r) | 95% CI | Sig. | R² | ICC | 95% CI | Sig. | |
| Non-contact forehead (Tno-c) | 0.47 | 0.36;0.57 | p < 0.001 | 0.22 | 0.3445 | −0.18; 0.63 | p < 0.001 |
| Ingestible capsule (Tcap) | 0.85 | 0.81;0.98 | p < 0.001 | 0.72 | 0.8730 | 0.55; 0.94 | p < 0.001 |
| eTact (Tetac) | 0.42 | 0.30;0.52 | p < 0.001 | 0.17 | 0.1229 | −0.09; 0.37 | p < 0.001 |