Table 3.

Results of antibody detection in clinical serum samples using the ICTS and the FAVN testa.

FAVN test
≥0.5 IU/ml <0.5 IU/ml Negative Total
ICTS Positive 331 7 0 338
Negative 35 245 138 418
Total 366 252 138 756
a

Relative to FAVN test, specificity of ICTS: (245 + 138)/(252 + 138) = 98.2%; sensitivity of ICTS: 331/(331 + 35) = 90.4%; Po = (383 + 331)/756 = 94.4%; Pe = [(366/756) × (338/756) + (390/756) × (418/756)] = 0.502; kappa = (Po − Pe)/(1 − Pe) = 0.4424/0.498 = 0.888.