Reproducibility assay of threshold cycle value quantified by RT-MRT-PCR from the standard plasmid.

Plasmid Concentrationa (viral copy/μl) Intra-assay
Inter-assay
Mean S.D. CV (%) Mean S.D. CV (%)
Genotype 1 101 39.25 0.60 1.54 39.12 0.17 0.44
104 27.56 0.13 0.46 27.41 0.26 0.93
107 17.82 0.18 1.00 18.13 0.29 1.60



Genotype 2 101 38.25 0.31 0.82 38.33 0.45 1.17
104 27.02 0.15 0.56 26.98 0.11 0.40
107 17.41 0.36 2.08 17.96 0.32 1.76



Genotype 3 101 38.33 0.61 1.60 37.50 0.37 0.99
104 26.69 0.05 0.19 26.62 0.08 0.31
107 16.86 0.37 2.17 17.49 0.26 1.48
a