Skip to main content
. Author manuscript; available in PMC: 2021 Apr 1.
Published in final edited form as: Am J Ophthalmol. 2019 Nov 23;212:57–68. doi: 10.1016/j.ajo.2019.11.018

Table 3.

Global Agreement and Reproducibility for Visual Field & Optical Coherence Tomographic Angiography

NFLP_MD v. VF_MD VF_MD v. VF_MD* NFLP_MD v. NFLP_MD*
Intraclass Correlation Coefficient (95% Confidence Interval) 0.808 (0.709 to 0.876) 0.946 (0.900 to 0.971) 0.954 (0.910 to 0.977)
Pooled Standard Deviation (dB) 1.78 1.03 0.63

The agreement between visual field mean deviation (VF_MD) and retina nerve fiber layer plexus mean deviation (NFLP_MD) measured on the same visit were assessed. Both normal and glaucoma groups were used in this analysis.

*

The between-visit reproducibility of VF_MD and NFLP_MD were measured in the glaucoma group using data from 2 visits 6 months apart.