Table 6.

Sensitivity and specificity of PEDV RT-iiPCR, PDCoV RT-iiPCR, and duplex PEDV/PDCoV rRT-PCR as compared to the reference singleplex PEDV rRT-PCR and PDCoV rRT-PCR based on testing 170 clinical samples.

Reference PEDV rRT-PCR or reference PDCoV rRT-PCR
Positive Negative Total
(a) PEDV RT-iiPCR compared to the reference singleplex PEDV rRT-PCR
PEDV RT-iiPCR Positive 86 1 87
Negative 2 81 83
Total 88 82 170
Sensitivity: 97.73%; Specificity: 98.78%; Accuracy: 98.24%



(b) Duplex PEDV/PDCoV rRT-PCR compared to the reference singleplex PEDV rRT-PCR
Duplex PEDV/PDCoV rRT-PCR Positive 87 3 90
Negative 1 79 80
Total 88 82 170
Sensitivity: 98.86%; Specificity: 96.34%; Accuracy: 97.65%



(c) PDCoV RT-iiPCR compared to the reference singleplex PDCoV rRT-PCR
PDCoV RT-iiPCR Positive 82 0 82
Negative 0 88 88
Total 82 88 170
Sensitivity: 100%; Specificity: 100%; Accuracy: 100%



(d) Duplex PEDV/PDCoV rRT-PCR compared to the reference singleplex PDCoV rRT-PCR
Duplex PEDV/PDCoV rRT-PCR Positive 79 0 79
Negative 3 88 91
Total 82 88 170
Sensitivity: 96.34%; Specificity: 100%; Accuracy: 98.24%

Notes: (1) Nucleic acids were extracted from all samples using the MagMAXTM Pathogen RNA/DNA Kit and Kingfisher-96 instrument from Thermo Fisher Scientific. (2) The reference PEDV rRT-PCR and PDCoV rRT-PCR: Ct < 40 positive. (3) The duplex PEDV/PDCoV rRT-PCR: Ct < 40 positive; Ct ≥ 40 negative.