Table 6.
Sensitivity and specificity of PEDV RT-iiPCR, PDCoV RT-iiPCR, and duplex PEDV/PDCoV rRT-PCR as compared to the reference singleplex PEDV rRT-PCR and PDCoV rRT-PCR based on testing 170 clinical samples.
Reference PEDV rRT-PCR or reference PDCoV rRT-PCR |
||||
---|---|---|---|---|
Positive | Negative | Total | ||
(a) PEDV RT-iiPCR compared to the reference singleplex PEDV rRT-PCR | ||||
PEDV RT-iiPCR | Positive | 86 | 1 | 87 |
Negative | 2 | 81 | 83 | |
Total | 88 | 82 | 170 | |
Sensitivity: 97.73%; Specificity: 98.78%; Accuracy: 98.24% | ||||
(b) Duplex PEDV/PDCoV rRT-PCR compared to the reference singleplex PEDV rRT-PCR | ||||
Duplex PEDV/PDCoV rRT-PCR | Positive | 87 | 3 | 90 |
Negative | 1 | 79 | 80 | |
Total | 88 | 82 | 170 | |
Sensitivity: 98.86%; Specificity: 96.34%; Accuracy: 97.65% | ||||
(c) PDCoV RT-iiPCR compared to the reference singleplex PDCoV rRT-PCR | ||||
PDCoV RT-iiPCR | Positive | 82 | 0 | 82 |
Negative | 0 | 88 | 88 | |
Total | 82 | 88 | 170 | |
Sensitivity: 100%; Specificity: 100%; Accuracy: 100% | ||||
(d) Duplex PEDV/PDCoV rRT-PCR compared to the reference singleplex PDCoV rRT-PCR | ||||
Duplex PEDV/PDCoV rRT-PCR | Positive | 79 | 0 | 79 |
Negative | 3 | 88 | 91 | |
Total | 82 | 88 | 170 | |
Sensitivity: 96.34%; Specificity: 100%; Accuracy: 98.24% |
Notes: (1) Nucleic acids were extracted from all samples using the MagMAXTM Pathogen RNA/DNA Kit and Kingfisher-96 instrument from Thermo Fisher Scientific. (2) The reference PEDV rRT-PCR and PDCoV rRT-PCR: Ct < 40 positive. (3) The duplex PEDV/PDCoV rRT-PCR: Ct < 40 positive; Ct ≥ 40 negative.