Table 8.

Sensitivity and specificity of PEDV RT-iiPCR and PDCoV RT-iiPCR as compared to the reference singleplex PEDV rRT-PCR and PDCoV rRT-PCR based on testing 170 clinical samples with different nucleic acid extraction methods.

(a) Comparison of PEDV RT-iiPCR and the reference singleplex PEDV rRT-PCR
MagMAX™ extraction and reference PEDV rRT-PCR
Positive Negative Total
taco™ mini extraction and PEDV RT-iiPCR Positive 83 0 83
Negative 5 82 87
Total 88 82 170
Sensitivity: 94.32%; Specificity: 100%; Accuracy: 97.06%
(b) Comparison of PDCoV RT-iiPCR and the reference singleplex PDCoV rRT-PCR
MagMAX™ extraction and reference PDCoV rRT-PCR
Positive Negative Total
taco™ mini extraction and PDCoV RT-iiPCR Positive 82 0 82
Negative 0 88 88
Total 82 88 170
Sensitivity: 100%; Specificity: 100%; Accuracy: 100%