Skip to main content
. 2018 May 12;258:41–48. doi: 10.1016/j.jviromet.2018.05.006

Table 2.

Sensitivity of QProbe RT-LAMP assays.

Copies/reaction 2000 200 20 2 0.2 Sensitivity (copies/reaction) Time required (h)
Real-time RT-PCR Positive/Number 
 upE 6/6 6/6 5/6 1/6 0/6 6.3 2
 ORF1a 6/6 6/6 4/6 0/6 0/6 13.6 2



Copies/reaction 2000 200 20 2 0.2
RT-LAMP Positive/Number
 N (turbidity) 6/6 6/6 2/6 1/6 0/6 20 0.5
 N (QP) 8/8 8/8 3/8 1/8 0/8 20 0.5
 ORF1a (QP) 8/8 8/8 5/8 0/8 0/8 15 0.5

QP: Quenching probe.