Skip to main content
. 2019 Mar 1;267:53–58. doi: 10.1016/j.jviromet.2019.02.010

Table 5.

Performance of novel QPrimer-based rRT-LAMP assay for IV and RSV compared with the reference rRT-PCR assay.

Target LAMP result rRT-PCR result
LAMP sensitivity % (95% CI) LAMP specificity % (95% CI)
Positive Negative
IAV Positive 26 0 89.7 (72.7–97.8) 100 (95.7–100)
Negative 3 84
IBV Positive 10 0 90.9 (58.7–99.8) 100 (96.5–100)
Negative 1 102
A/H1pdm Positive 16 0 88.9 (65.3–98.6) 100 (96.2–100)
Negative 2 95
A/H3 Positive 10 0 90.9 (58.7–99.8) 100 (96.5–100)
Negative 1 102
RSV A Positive 17 0 85.0 (62.1–96.8) 100 (96.1–100)
Negative 3 93
RSV B Positive 13 0 92.9 (66.1–99.8) 100 (96.3–100)
Negative 1 99