Table 5.

Limits of detection of each target used in this study are expressed as the minimum number of target gene copies in 25 μl PCR volume needed to produce a band visible by gel electrophoresis (detection limit of multiplex PCR) or to produce a reactivity of >2 PN ratio by electronic microarray (detection limit in electronic microarray).

Target pathogen Target gene Detection limit by multiplex PCR/RT-PCR Detection limit by electronic microarray
BRD virus multiplex RT-PCR
BoHV-1 gB 100 1000
BRSV G 1 1
BPI-3 M 1 10
BVDV 1 5’UTR 1 1
BVDV 2 5’UTR 1 10
BRD bacteria multiplex PCR
H. somni 16S rDNA 1 1
P. multocida ulaR 1 1
M. bovis 16S rDNA 1 1
M. haemolytica nmaA 10 10
tbpB 10 100
lktA 100 1000
BED virus multiplex RT-PCR
BCoV N 1 1
BRoV A VP6 1 1
BRoV C VP6 100 1000
BRoV B VP6 1 10
BToV (common) N 1000 10000
BToV (Breda) N 1000 1000
BVDV 1 5’UTR 1 1
BVDV 2 5’UTR 1 10
BED bacteria multiplex PCR
C. perfringens plC 10 10
E. zuernii ITS1 10 100
E. bovis ITS1 10 10
C. parvum COWP 100 1000
S. e. Dublin A1118 10 100
S. e. Typhimurium STM4493 100 100
E. coli eaeA 10 10
stx-1 10 100
Lt 1 10
K99 10 100