Skip to main content
. 2017 Dec 15;253:18–25. doi: 10.1016/j.jviromet.2017.12.005

Table 3.

Limit of detection of RT-LAMP, conventional one-step RT-PCR and real-time RT-PCR assay using virus titre as plaque forming units (PFU).

Influenza strain Target gene RT-LAMP (PFU/reaction) Conventional RT-PCR (PFU/reaction) WHO rRT-PCR (PFU/reaction)
A/California/7/2009 (H1N1) pdm09-like virus M 0.01 0.1 0.001
A/California/7/2009 (H1N1) pdm09-like virus HA 0.01 1 0.01
A/Texas/50/2012 (H3N2)-like virus HA 0.1 0.1 0.1
A/Malaysia/2089302/2009 (H1N1)-like virus HA 0.1 1 0.01