. 2015 May 28;43(6):606–611. doi: 10.1016/j.ajic.2015.02.025

Table 2.

Prevalence of cross-contamination and viral loads of sites after EMS calls

Sites No decontamination Current practices H2O2 intervention
EPCR keyboard 3/5 54.2 ± 91 1/5 5.8 ± 11 1/5 72 ± 158
EPCR touchscreen 1/5 67.4 ± 138 3/5 76.4 ± 84 2/5 90 ± 141
Glucometer 2/5 8.40 ± 7 2/5 10.4 ± 10 1/5 15 ± 26
Headphones 3/3 128 ± 188 1/3 7.33 ± 11 1/3 9 ± 14
Inside cab 3/5 390 ± 15 2/5 7.60 ± 9 5/5 96 ± 129
Jumpbag handle 4/5 40.2 ± 28 5/5 226 ± 153 3/5 230 ± 315
LIFEPAK keypad 3/5 39.4 ± 58 3/5 58.4 ± 70 1/5 33 ± 68
MDT keyboard 3/5 92.4 ± 173 5/5 33.0 ± 9 0/5 <10 ± 0
MDT touchscreen 2/5 13.4 ± 14 1/5 10.4 ± 12 2/5 24 ± 35
Portable radios 3/5 10.4 ± 191 3/5 11.4 ± 10 3/5 20 ± 17

NOTE. Values are positive samples/total samples or mean ± SD.

EPCR, electronic patient care record; MDT, mobile data terminal.