Table 1.

Limits of detection (LoD95) for respiratory targets

Viral organism LoD95 (pfu/mL)
 Adenovirus type 3 79
 Adenovirus type 4 71
 Human bocavirus 86 (cop/μL)
 Coxsackievirus A 1 × 103
 Coxsackievirus B 1 × 102
 Echovirus 1
 Influenza A H1N1-09 7
 Coronavirus 229E 6
 Coronavirus HKU1 87 (cop/μL)
 Coronavirus NL63 1
 Coronavirus OC43 7
 Influenza A H3N2 1
 Influenza B 1
 Parainfluenza virus type 1 10
 Parainfluenza virus type 2 6
 Parainfluenza virus type 3 100
 Parainfluenza virus type 4A 88
 Rhinovirus 1
 Respiratory syncytial virus A 1
 Respiratory syncytial virus B 1
 Human metapneumovirus A 562
 Human metapneumovirus B
316
Bacterial organism
LoD95 (cfu/mL)
Acinetobacter baumannii 7.5 × 104
Chlamydophila pneumoniae 100
Haemophilus influenzae (nonspecific: types A-F) 1 × 103-1 × 104
H influenzae type B 1 × 103-1 × 104
Klebsiella pneumoniae 1 × 103
Legionella pneumophila 750
Mycoplasma pneumoniae 10
Moraxella catarrhalis 100
Neisseria meningitidis 72
Pseudomonas aeruginosa 1 × 103
Bordetella pertussis 6.8 × 104
Staphylococcus aureus 100
Streptococcus pneumoniae 464
Str pyogenes 1 × 104
Panton-Valentine leukocidin (cytotoxin) 3.20 × 104
Methicillin resistance (antibiotic resistance) 100

NOTE. Analytical sensitivities were determined following Clinical and Laboratory Standards Institute guidelines. At the concentrations listed, the corresponding genetic target is a true positive result 95% of the time. The Supplemental Data describes the analytical validation procedures, including the measurement of >95% test specificity.

cfu, colony forming units; cop, copies; pfu, plaque forming units; LoD95, 95% Limit of Detection.