Skip to main content
. Author manuscript; available in PMC: 2020 Dec 1.
Published in final edited form as: Hear Res. 2020 Apr 10;391:107969. doi: 10.1016/j.heares.2020.107969

Table 3.

Increase in charge of the global C-level needed to achieve the same loudness level for the TIPS conditions relative to the CIS reference for each participant, both in nano Coulomb (nC) and as a percentage (%).

Subject TIPS 25% TIPS 50% TIPS 75%
nC % nC % nC %
S1 0.02 1.82 0.10 9.45 0.20 17.65
S2 0.02 1.82 0.07 7.49 0.11 11.44
S3 0.00 0.00 0.32 24.20 0.62 46.12
S4 0.05 1.82 0.20 7.49 0.26 9.45
S5 0.02 1.82 0.05 5.57 0.18 19.79
S6 0.09 3.68 0.18 7.49 0.38 15.54
S7 0.09 3.68 0.14 5.57 0.33 13.48
S8 0.00 0.00 0.12 7.49 0.22 13.48
mean 0.036 1.830 0.147 9.343 0.287 18.368