Skip to main content
. 2013 Jul 25;193(2):492–497. doi: 10.1016/j.jviromet.2013.07.027

Table 1.

Limit of detection of the ExiStation HBV diagnostic system.

NIBSC standard HBV DNA concentration (IU/mL) Positive No. (%) of replicates (n = 24)
1000 24 (100)
333.3 24 (100)
111.1 24 (100)
37.0 24 (100)
12.3 23 (95.8)
4.1 20 (83.3)
1.03 10 (41.6)
0.52 5 (20.8)

Detection limit by probit analysis: 9.55 IU/mL (95% confidence interval: 4.68–19.50 IU/mL).