Repeatability and reproducibility of ExiStation HBV diagnostic system.

Mean (log10 IU/mL) Within-run SD (log10 IU/mL) Within-run CV (%) Between-day SD (log10 IU/mL) Between-day CV (%) Total SD (log10 IU/mL) Total CV (%)
Not detected
1.92 0.18 9.4 0.14 7.1 0.22 11.8
3.88 0.08 2.1 0.11 2.9 0.14 3.6
6.84 0.07 1.1 0.09 1.3 0.12 1.7