Table 3.
Intra-and inter-assay reproducibility of real-time RT-PCR
Concentration of specimens (TCID50/ml) | Replicate (Cta value) |
Ct mean (S.D.b) | %CVc | |||
---|---|---|---|---|---|---|
1 | 2 | 3 | 4 | |||
Intra-assay | ||||||
105 | 15.68 | 15.54 | 15.41 | 15.38 | 15.50 (0.14) | 0.9 |
104 | 19.01 | 18.54 | 18.98 | 18.72 | 18.81 (0.23) | 1.2 |
103 | 21.23 | 22.43 | 22.12 | 21.94 | 21.93 (0.51) | 2.3 |
102 | 25.23 | 25.36 | 26.02 | 25.43 | 25.51 (0.35) | 1.4 |
101 | 28.56 | 28.39 | 28.12 | 28.40 | 28.37 (0.17) | 0.6 |
100 | 31.98 | 31.77 | 32.12 | 31.95 | 31.96 (0.16) | 0.5 |
Inter-assay | ||||||
105 | 14.98 | 15.72 | 15.47 | 15.84 | 15.50 (0.39) | 2.5 |
104 | 18.89 | 18.62 | 18.63 | 18.33 | 18.62 (0.22) | 1.2 |
103 | 22.31 | 22.58 | 22.21 | 22.02 | 22.28 (0.22) | 1.0 |
102 | 25.13 | 25.46 | 25.71 | 24.98 | 25.32 (0.33) | 1.3 |
101 | 28.10 | 28.64 | 28.01 | 28.34 | 28.27 (0.25) | 0.9 |
100 | 31.25 | 31.76 | 32.53 | 31.89 | 31.86 (0.54) | 1.7 |
Cycle threshold value.
Standard deviation.
Coefficient of variation = S.D./Ct mean.