Table 2.
Fluence of 1H standard implant (×1015 H cm−2).a
| Measurement | 1H/16O | 1H/31P | 16O1H/16O2 |
|---|---|---|---|
| Crystal Lode #1 | 5.57 | 5.50 | 5.35 |
| Crystal Lode #2 | 5.54 | 5.59 | 5.53 |
| Mean Crystal Lode | 5.56 ± 0.04b | 5.55 ± 0.13b | 5.44 ± 0.26b |
| Lake Baikal #1 | 5.88c | 5.92c | 5.13c |
| Lake Baikal #2 | 5.95 | 5.99 | 5.52 |
| Lake Baikal #3 | 5.81 | 5.84 | 5.64 |
| Mean Lake Baikal | 5.88 ± 0.14b | 5.92 ± 0.15b | 5.43 ± 0.53b |
| Mean from atom ratiosd | 5.73 ± 0.41e |
Nominal implant fluence 6 × 1015 H cm−2.
Uncertainties on mean values are 2σ SDs of the measurements. Statistical uncertainties are insignificant.
Measured at 300 pAmps beam current; others were measured at 500 pAmps.
Mean of results from atomic ratios (each standard and ratio given the same weight).
Uncertainty does not include the uncertainty in the true water contents of the two apatites.