Skip to main content
. 2020 Apr 3;15(4):e0230453. doi: 10.1371/journal.pone.0230453

Table 3. Diagnostic performance of Visitect CD4 LFA at different CD4 cut-offs of the reference test.

CD4 range (n; %) Sensitivity (%), 95% CI Specificity (%), 95% CI
<100 (108; 15.3%) 98.1 [93.5–99.8] 68.0 [64.1–71.7]
100–200 (113; 16.0%) 92.0 [85.4–96.3] 67.4 [63.5–71.2]
<200 (221; 31.2%) 95.0 [91.3–97.5] 81.9 [78.2–85.2]
<350 (417; 58.9%) 68.8 [64.1–76.7] 96.2 [93.3–98.1]