Table 3. Diagnostic performance of Visitect CD4 LFA at different CD4 cut-offs of the reference test.
CD4 range (n; %) | Sensitivity (%), 95% CI | Specificity (%), 95% CI |
---|---|---|
<100 (108; 15.3%) | 98.1 [93.5–99.8] | 68.0 [64.1–71.7] |
100–200 (113; 16.0%) | 92.0 [85.4–96.3] | 67.4 [63.5–71.2] |
<200 (221; 31.2%) | 95.0 [91.3–97.5] | 81.9 [78.2–85.2] |
<350 (417; 58.9%) | 68.8 [64.1–76.7] | 96.2 [93.3–98.1] |