Table 2.
Experimental gem values for 5.5 mM Eu{(+)-facam}3 in anhydrous DMSO versus those obtained by reported by Sánchez-Carnerero et al.5.
Transition (wavelength) | gem (Sánchez-Carnerero et al.) | gem (SM-CPL) | gem (SS-CPL) |
---|---|---|---|
5D0→7F1 (595.2 nm) | −0.78 ± 0.01 (1.3%)a | −0.70 ± 0.06 (9%)b | −0.763 ± 0.018 (2.4%)c |
5D0→7F2 (613.5 nm) | +0.07 ± 0.01 (14%)a | +0.07 ± 0.01 (14%)a | +0.114 ± 0.018 (16%)c |
aNominal SM-CPL instrument uncertainty (assumed).
bUncertainty calculated from SNR estimated by fitting Gaussian curves to CPL and total intensity measurements using least mean squares regression.