Skip to main content
. 2020 Apr 3;10:5910. doi: 10.1038/s41598-020-62474-1

Table 2.

C6Mα3 ELISA Technical Validation Data.

Technical validation C6Mα3
IC50 1.34 ng/mL
Detection range 0.219–8.6 ng/mL
Intra-assay variationa 8%
Inter-assay variationa 14%
Dilution recovery in seruma 96%
Interference biotin, low/higha 98%/105%
Interference lipemia, low/higha 123%/96%
Interference hemoglobin, low/higha 109%/104%
Freeze-thaw stabilitya 105%
Analyte stabilitya 87%
Spiking recoverya 54%

aMean recovery percentages are reported.