Skip to main content
. 2020 Feb 8;209(2):139–150. doi: 10.1007/s00430-019-00655-0

Table 7.

Lack of detection of four HIV-2 strains and two HIV-2 primary isolates by the new Alere RDT compared to quantitative PCR

PCR New RDT Virus strain/isolate HIV-2 group
(c/mL) Antibody Ag
205,000 CBL-20 A
450,000 Rod10 A
805,000 EHO B
1,000,000* ATU A
1,000,000* MVP1845852 A
1,000,000* MVP1847457 A
0/6 positive (0%)