Table 4.
Near‐patient assay/manufacturer | No. of samples available | Median concentration of influenza A virus RNA (copies/mL) | p‐valuea | ROC curve cut‐off for 100% sensitivity (copies/mL) | |
---|---|---|---|---|---|
Negative | Positive | ||||
Binax Now Influenza A&B/Binax Inc. | 28 | 1.7 × 103 | 5.1 × 104 | 0.002 | 2.7 × 103 |
Quick S‐Influ A/B/Seiken | 24 | 1.9 × 103 | 8.0 × 104 | 0.048 | 2.6 × 103 |
Influ‐A&B Respi‐Strip/Coris Bioconcept | 22 | 2.1 × 103 | 7.2 × 104 | 0.014 | 9.3 × 102 |
Actim Influenza A&B/Medix Biochemica | 32 | 3.4 × 103 | 3.1 × 104 | 0.004 | 6.8 × 102 |
ROC, receiver operating characteristic.
aMann–Whitney U‐test.