Table 1.
Characteristics of BioFire FilmArray RP, Nanosphere Verigene RV+ Test and Hologic Gen-Probe Prodesse assays
| Name | BioFire FilmArray | Verigene | GenProbe Prodesse |
|---|---|---|---|
| Technology | Melting curve analysis | Gold nanoparticles with silver signal amplification | Melting curve analysis |
| Assays | Respiratory panel | Respiratory virus plus test | ProFlu+, ProFAST+, ProAdeno+, ProParaflu+, Pro hMPV+ |
| Targets |
|
|
|
| Throughput | 1 sample per instrument | 1 sample per processor | 14 samples per run |
| Run time (hours) | 1 | <2.5 | 4–5 |
| Hands-on time | 2 minutes | 5 minutes | 1.5 hours |
| Sample preparation included? | Yes | Yes | No |
| Reagent storage conditions | Room temperature | 2–8°C and −20°C | −70°C |
FluA, influenza A virus; hMPV, human metapneumovirus; RSV, respiratory syncytial virus.