. 2004 Jul 7;31(2):116–122. doi: 10.1016/j.jcv.2004.04.003

Table 2.

Comparison of the diagnostic efficacies of the Ad-specific PCR methods with that of virus culture or immunofluorescence in 362 nasal aspirates samples

Diagnostic Method PCR Ad consensus
PCR Ad hexon
Positive Negative Positive Negative
Virus culture/Immunofluorescence
Positive 95 2 94 3
Negative 18a 247 13 252


Total 

362 

362


Sensitivity (%) 

97.9 

96.9
Specificity (%) 93.2 95
PPV (%) 84 87.8
NPV (%) 99.1 98.8
a

Thirteen specimens were positive in both PCR Ad consensus and PCR Ad hexon but not in according to the culture/immunofluorescence method. Five additional samples were only positive according to PCR Ad consensus, these samples were re-extracted and controlled as positive.