Skip to main content
. 2020 Apr 2;33(2):83–98. doi: 10.1089/jamp.2019.1540

Table 7.

Aerosolization and Lung Delivery Efficiency for Oral Administration (Case 4 Device, MP 2) Through the 5-Year-Old Pediatric Mouth–Throat Geometry

Description MP 2
Chamber (%) 11.2 (2.1)
Device (%) 10.6 (1.5)
MP (%) 4.6 (0.5)
MP ED (%) 73.6 (0.5)
MT loss (%) 6.6 (1.1)
Tracheal filter (%) 63.8 (2.2)
Recovery (%) 96.8 (3.1)

Mean aerosol characteristics with SD shown in parenthesis, n = 3.

Device was operated with ambient air and ventilation bag.

MT, mouth–throat.