Table 1:

Imaging parameters for the sagittal fat-suppressed T1-weighted spoiled gradient-echo (T1-SPGR) sequence in the SKI10 image dataset and the two clinical MR sequences including sagittal fat-suppressed T2-weighted fast spin-echo (T2-FSE) and sagittal proton density-weighted fast spin-echo (PD-FSE).

Sequence T2-FSE PD-FSE T1-SPGR
Field Strength (T) 3.0 3.0 1, 1.5, 3.0
TR [ms] 4680 1900 Unavailable
TE [ms] 80 19 Unavailable
Flip Angle [degree] 90 90 Unavailable
Pixel Bandwidth [Hz] 163 122 Unavailable
Echo Train Length 20 4 N/A
Field of View [cm] 14 14 14
Slice Thickness [mm] 3 2 1
Slice Gap [mm] 0.5 0.5 N/A
Number of Slices 23~30 32~42 90~120
Matrix Size 512×512 512×512 360×300

Approximate values as actual matrix size and field of view vary with the size of the knee joint in the SKI10 image dataset