. 2020 Mar 6;92(7):5276–5285. doi: 10.1021/acs.analchem.9b05836

Table 2. Results for Ion Readings (mM) and pH with (w/) and without (w/o) the Novel Algorithm.

      K+ Na+ Ca2+ pH
Sol 1 concentration   2.7 50 5 5.37
  w/ algorithm chip 2.68 54.3 2.99 5.36
    err. (%) 0.61 8.7 40 0.25
  w/o algorithm chip 2.35 57.8 2.69 4.76
    err. (%) 13 16 46 11.4
Sol 2 concentration   9 147 8 5.47
  w/ algorithm chip 8.58 148 6.4 5.47
    err. (%) 4.7 0.85 20 0.054
  w/o algorithm chip 8.33 150 5.86 5.52
    err. (%) 7.5 1.9 27 0.97
Sol 3 concentration   20 100 1.2 5.51
  w/ algorithm chip 18.8 96.4 1.23 5.36
    err. (%) 6.2 3.6 2.9 2.7
  w/o algorithm chip 17.9 92.5 1.02 5.53
    err. (%) 11 7.5 15 0.35
Sol 4 concentration   8.54 46.49 3.79 5.85
  w/ algorithm chip 8.82 49 2.51 5.6
    err. (%) 3.3 5.4 34 4.3
  w/o algorithm chip 7.75 49.4 2.08 4.89
    err. (%) 9.3 6.3 45 16.5