Skip to main content
. 2020 Mar 24;20(6):1800. doi: 10.3390/s20061800

Table 5.

Results of AA determination on the proposed Au-gr/CVE sensor and potentiometric titration in fruit juices (n = 3, P = 0.95).

Samples Found Using Au-gr/CVE RSD, % Found by Potentiometric Titration RSD, % F-Test t-Test
Cherry-apple juice 0.81 ± 0.06 6.8 0.76 ± 0.07 8.7 1.43 0.67
Apple juice for children 0.47 ± 0.01 2.2 0.42 ± 0.03 7.6 9.00 1.82
Apple juice 0.16 ± 0.01 5.0 0.16 ± 0.01 4.5 1.18 0.49
Apple nectar clarified 0.09 ± 0.01 6.5 0.09 ± 0.01 9.0 1.97 0.19

Fcrit. = 19.00 at P = 0.95 и df1 = 2, df2 = 2; tcrit. = 2.78 at P = 0.95 и df = 4.